Product Description
Product
Name
C-reactive
Protein(CRP)Test Kit (Immunofluorescence
Chromatography)
Â
Purpose
To detect the content
of CRP in human plasma/serum/whole blood in vitro
quantitatively.For Professional Use only.
Â
Principles of
Detection
The CRP Test is a
fluorescent lateral flow immunoassay. The test consists of: 1) a
conjugate pad containing CRP antibody conjugated with
fluorescent microspheres and a
control antibody (goat anti-rabbit IgG antibody) conjugated with
fluorescence microspheres, 2) a nitrocellulose membrane strip
containing a test line (T line)
and a control line (C line). The T line is pre-coated with another
CRP antibody, and the C line is pre-coated with a control line
antibody (rabbit IgG
antibody).
Â
CRP concentration in
specimen is positively correlated with the fluorescent density of
the microspheres, after the measure and calculation of the
quantitative fluorescence
immunoassay analyzer, then the analyzer would calculate the
concentration of CRP according to the preset calibration curve in
the machine and display the
result, detection unit is mg/L. C line should appear in any cases,
which indicates the reaction system is effective.
Â
Storage conditions and
shelf life
Store in a sealed
aluminum foil bag at 2~*0°C. After unpacking the aluminum foil bag
under temperature *0~*0 ºC , humidity *5~*5%, please use
it immediately.
Shelf life *8
months.
Â
Applicable
instrument
Avid fluorescent
immunoanalyzer model: AI-IF **0
Manufacturer&
After-Sales Company: Shenzhen Aivd Biotechnology Co.,
LTD.
Address:C**1,Building
B5,China Merchants Guangming Science Park, #***9 Guanguang Road,
Guangming District, Shenzhen,Guangdong Province,
China
Tel: ************2
Fax:************5
Â
Specimen collection
and preparation
1. Human
plasma/serum/whole blood samples can be used for this
test.
2. Specimen
collection:
Plasma/Serum:
The serum or plasma
should be separated by centrifugation immediately after the
collection of whole blood. Serum specimens must be collected in
standard or vacuum test tubes
with serum separating gel.
Specimens can be
stored at **8°C for 7 days after processing. The specimens frozen
at **0°C shouldn't be over 1 month. Multiple freeze-thaw
cycles shouldn't over 3
times.
Whole
Blood:
Collect blood specimen
into a collection tube containing EDTA or heparin or citric acid,
evenly mixed before use. Do not use hemolyzed blood for
testing.
Whole blood specimens
should be stored at **8°C for no more than 4 hours if not test
immediately.
3. Do not test
patients' samples with severe hemolysis, lipemia and jaundice, all
specimens should be treated as infectant.
4. Restore the
specimen to ****0°C before the detection, cryopreserved specimen
should be completely thawed, rewarmed and evenly mixed before
use.
Â
Test
procedure
Read the complete
instruction before the test
1. Bring
the specimen and test
components to room
temperature (recommended temperature is about ****0 ºC ) to
test. Open the aluminum foil
bag containing
the test kit after it get back to the room temperature. Keep
the test kit in a dry place to prevent spoilage from dampness.
Before unpacking, check whether the
aluminum foil bag is in good condition. Do not use it if it is
damaged. The test card should be used within 1 hour after
unpack.
2. Open the
instrument as its instruction and Insect the ID card to get the
curve.
3. Collect
5μl of
sample(plasma/serum/whole blood), add to the sample diluent
(mix thoroughly for *0
sec with vortex mixer)
collect *0μl of
a sample mixture and load it into
the sample vertically and slowly.
4. Leave the
sample-loaded cartridge at room temperature (recommended
temperature is about ****0ºC) for 5 minutes. After time is
over, immediately insert it into the cartridge
holder of the instrument, press "Test" and read the
results.
5. Treat the
tested card as potential biological hazard
6. Quality
Control:
1) As an
effectiveness indicator, C line should appear in any
cases.
2) Aivd's internal
control materials or other approved applicable control
materials can be used to control the quality of the product,
the test results should be within the
specified quality control range.
Note: Perform the
test under the condition of temperature *0~*0ºC, humidity
*5~*5%.
Â
Reference
range
Reference
values:<*0mg/L.
The reference
vallue
has been verified with reference to similar products already on
the market. Each laboratory can build own reference range
according to
actual
situation.
Country: |
China |
Model No: |
-
|
FOB Price: |
( Negotiable ) (Negotiable)
Get Latest Price
|
Place of Origin: |
- |
Price for Minimum Order: |
- |
Minimum Order Quantity: |
100 Piece |
Packaging Detail: |
- |
Delivery Time: |
- |
Supplying Ability: |
100000 Piece per Day |
Payment Type: |
T/T, L/C |
Product Group : |
- |